An investigation of input level range for the nucleus 24 cochlear implant system: speech perception performance, program preference, and loudness comfort ratings

Ear Hear. 2003 Apr;24(2):157-74. doi: 10.1097/01.AUD.0000058107.64929.D6.

Abstract

Objective: Cochlear implant recipients often have limited access to lower level speech sounds. In this study we evaluated the effects of varying the input range characteristics of the Nucleus 24 cochlear implant system on recognition of vowels, consonants, and sentences in noise and on listening in everyday life.

Design: Twelve subjects participated in the study that was divided into two parts. In Part 1 subjects used speech processor (Nucleus 24 SPrint trade mark ) programs adjusted for three input sensitivity settings: a standard or default microphone sensitivity setting (MS 8), a setting that increased the input sensitivity by 10.5 dB (MS 15), and the same setting that increased input sensitivity but also incorporated the automatic sensitivity control (ASC; i.e., MS 15A) that is designed to reduce the loudness of noise. The default instantaneous input dynamic range (IIDR) of 30 dB was used in these programs (i.e., base level of 4; BL 4). Subjects were tested using each sensitivity program with vowels and consonants presented at very low to casual conversational levels of 40 dB SPL and 55 dB SPL, respectively. They were also tested with sentences presented at a raised level of 65 dB SPL in multi-talker babble at individually determined signal to noise ratios. In addition, subjects were given experience outside of the laboratory for several weeks. They were asked to complete a questionnaire where they compared the programs in different listening situations as well as the loudness of environmental sounds, and state the setting they preferred overall. In Part 2 of the study, subjects used two programs. The first program was their preferred sensitivity program from Part 1 that had an IIDR of 30 dB (BL 4). Seven subjects used MS 8 and four used MS 15, and one used the noise reduction program MS 15A. The second program used the same microphone sensitivity but had the IIDR extended by an additional 8 to 10 dB (BL 1/0). These two programs were evaluated similarly in the speech laboratory and with take-home experience as in Part 1. RESULTS PART 1: Increasing the microphone input sensitivity by 10.5 dB (from MS 8 to MS 15) significantly improved the perception of vowels and consonants at 40 and 55 dB SPL. The group mean improvement in vowel scores was 25 percentage points at 40 dB SPL and 4 percentage points at 55 dB SPL. The group mean improvement for consonants was 23 percentage points at 40 dB SPL and 11 percentage points at 55 dB SPL. Increased input sensitivity did not significantly reduce the perception of sentences presented at 65 dB SPL in babble despite the fact that speech peaks were then within the compressed range above the SPrint processor's automatic gain control (AGC) knee-point. Although there was a demonstrable advantage for perception of low-level speech with the higher input sensitivity (MS 15 and 15A), seven of the 12 subjects preferred MS 8, four preferred MS 15 or 15A, and one had no preference overall. Approximately half the subjects preferred MS 8 across the 18 listening situations, whereas an average of two subjects preferred MS 15 or 15A. The increased microphone sensitivity of MS 15 substantially increased the loudness of environmental sounds. However, use of the ASC noise reduction setting with MS 15 reduced the loudness of environmental sounds to equal or below that for MS 8. RESULTS PART 2: The increased instantaneous input range gave some improvement (8 to 9 percentage points for the 40 dB SPL presentation level) in the perception of consonants. There was no statistically significant increase in vowel scores. Mean scores for sentences presented at 65 dB SPL in babble were significantly lower (5 percentage points) for the increased IIDR setting. Subjects had no preference for the increased IIDR over the default. The IIDR setting had no effect on the loudness of environmental sounds.

Conclusions: Given the fact that individuals differ in threshold (T) and comfort (C) levels for electrical stimulation, and preferred microphone sensitivity, volume control, and noise-reduction settings, it is essential for the clinicid recipient to determine what combination is best for the individual over several sessions. The results of this study clearly show the advantage of using higher microphone sensitivity settings than the default MS 8 to provide better speech recognition for low-level stimuli. However, it was also necessary to adjust other parameters such as map C levels, automatic sensitivity control and base level, to optimize loudness comfort in the diversity of listening situations an individual encounters in everyday life.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Auditory Threshold*
  • Cochlear Implantation / instrumentation*
  • Deafness / surgery*
  • Female
  • Humans
  • Male
  • Sensitivity and Specificity
  • Speech Perception*