Specific determination of bromate in bread by ion chromatography with ICP-MS

Shokuhin Eiseigaku Zasshi. 2002 Dec;43(6):348-51. doi: 10.3358/shokueishi.43.348.

Abstract

A sensitive method for detecting bromate in bread by ion chromatography with inductively-coupled plasma mass spectrometry (IC/ICP-MS) was developed. Bromate was extracted from bread with water. The clean-up procedure included a 0.2 micron filter, a C18 cartridge for defatting, a silver cartridge to remove halogen anions, a centrifugal ultrafiltration unit to remove proteins, and a cation-exchange cartridge to remove silver ions. A 500 microL sample solution was applied to IC/ICP-MS. The detection limit and the quantitation limit of bromate in the solution were 0.3 ng/mL and 1.0 ng/mL, expressed as HBrO3, respectively, which corresponded to 2 ng/g and 5 ng/g, respectively, in bread. Recovery of bromate was about 90%, and the CV was about 2%. Based on the detection limit in solution and recovery from bread, the detection limit of bromate in bread was estimated to be 2 ng/g.

MeSH terms

  • Bread / analysis*
  • Bromates / analysis*
  • Chromatography, Ion Exchange / methods
  • Mass Spectrometry / methods*
  • Sensitivity and Specificity