Dynamic exercise imaging with an MR-compatible stationary cycle within the general electric open magnet

Magn Reson Med. 2003 Mar;49(3):581-5. doi: 10.1002/mrm.10364.

Abstract

Many cases of muscular ischemia do not manifest without increased metabolic demand. Hence, diagnosis of intermittent claudication often requires inducing physiologic challenge, such as by exercise. Cine phase-contrast MRI can concurrently acquire cross-sectional vascular anatomy and through-plane blood velocities, enabling blood flow rate quantification. An MR-compatible stationary cycle was designed, constructed, and tested for flow quantification in large arteries during lower-limb exercise in a General Electric Signa SP 0.5 T open magnet. The cycle demonstrated smooth cycling during image acquisition, has freewheeling capability, is adjustable for subject size and strength, and can quantify workload. A healthy 59-year-old male was imaged at the supraceliac and infrarenal levels of the abdominal aorta at rest and during exercise. An exercise workload of 47.9 W was achieved. His heart rate increased from 52 to 78 bpm, supraceliac flow increased from 1.7 to 3.7 L/min, and infrarenal flow increased from 0.4 to 3.2 L/min from rest to exercise.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Aorta, Abdominal / physiology*
  • Blood Flow Velocity
  • Equipment Design
  • Exercise / physiology*
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Male
  • Middle Aged
  • Reference Values