[Measurement of left ventricular chamber and myocardial volume in hypertrophic cardiomyopathy patients by ECG-gated myocardial perfusion SPECT: application of a newly developed edge-detection algorithm]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2002 Dec;58(12):1586-91. doi: 10.6009/jjrt.kj00000921536.
[Article in Japanese]

Abstract

Quantitative gated SPECT (QGS) software has been reported to demonstrate inaccurate edge detection in the left ventricular chamber in hypertrophic cardiomyopathy patients. In this study we developed a method to calculate left ventricular volume (LVV) and left myocardial volume (LMV) from gated SPECT data using a newly developed edge-detection algorithm, and we compared it with the QGS method of calculating LVV and LMV in a phantom study. Our method gave more accurate measurements LVV and LMV whereas the QGS method underestimated LMV. Compared with QGS LVV and LMV, our method yielded better results in the phantom study.

MeSH terms

  • Algorithms
  • Cardiomyopathy, Hypertrophic / diagnostic imaging*
  • Cardiomyopathy, Hypertrophic / physiopathology
  • Gated Blood-Pool Imaging
  • Humans
  • Image Processing, Computer-Assisted
  • Phantoms, Imaging*
  • Software
  • Stroke Volume*
  • Tomography, Emission-Computed, Single-Photon*
  • Ventricular Function, Left*