Comparison of procedures for evaluating laboratory performance in external quality assessment schemes for lead in blood and aluminum in serum demonstrates the need for common quality specifications

Clin Chem. 2002 Nov;48(11):2000-7.

Abstract

Background: The different scoring methods used by eight European External Quality Assessment Schemes (EQASs) for occupational and environmental laboratory medicine were compared to develop suitable quality specifications as a step toward harmonization.

Methods: Real results for blood lead and serum aluminum assays, reported by participants in Italian and United Kingdom EQASs, were evaluated according to individual scheme scoring criteria. The same results were then used to produce z scores using scheme-based between-laboratory SDs as the estimate of variability to determine whether simple performance-derived quality specifications produced better agreement among schemes.

Results: The schemes gave conflicting assessments of participants' performance, and participants judged to be successful by one scheme could be defined as performing inadequately by another. An approach proposed by Kenny et al. (Scand J Clin Lab Invest 1999;59:585), which uses clinical inputs to set targets for analytical imprecision, bias, and total error allowable, was then used to elaborate quality specifications.

Conclusions: We suggest that the CLIA '88 recommendations for blood lead (+/- 40 micro g/L or +/- 10% of the target concentration, whichever is the greater) could be used as a quality specification, although a revision to +/- 30 micro g/L or +/- 10% is recommended. For serum aluminum, a suitable quality specification of +/- 5 micro g/L or +/- 20% of the target concentration, whichever is the greater, is suggested. These specifications may be used to compare laboratory performance across schemes.

Publication types

  • Comparative Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aluminum / blood*
  • Clinical Laboratory Techniques / standards*
  • Environmental Pollutants / blood*
  • Environmental Pollutants / standards
  • European Union
  • Humans
  • Lead / blood*
  • Occupational Exposure
  • Quality Control
  • Reference Values

Substances

  • Environmental Pollutants
  • Lead
  • Aluminum