Corneal impact of ultrasound and bevel position in phacoemulsification

J Cataract Refract Surg. 2002 Sep;28(9):1667-70. doi: 10.1016/s0886-3350(02)01306-8.

Abstract

Purpose: To investigate the amplitude of ultrasound waves that reach the cornea with the phacoemulsification tip at different angles.

Setting: Institute for Applied Mechanics and Control Engineering, University of Siegen, Siegen, Germany.

Methods: A model eye was constructed with an ultrasound tip (angle 45 degrees) in the typical position during the procedure and a sensor mounted in the apex of the artificial cornea. The sensor measured the incoming ultrasound waves. Thirty bursts of ultrasound energy were set at 100% power with the bevel of the tip facing up toward the cornea, and 30 bursts were set with the bevel facing down toward the lens.

Results: There was no statistically significant difference in the measurements with the tips bevel up and bevel down.

Conclusions: There was no significant difference in the corneal load from ultrasound waves with the phacoemulsification tip in a bevel-up or bevel-down position.

Publication types

  • Comparative Study

MeSH terms

  • Cornea / radiation effects*
  • Humans
  • Models, Biological
  • Phacoemulsification / methods*
  • Ultrasonic Therapy*