Pathway selection by pulsed field gradients

Magn Reson Med. 2002 Sep;48(3):540-2. doi: 10.1002/mrm.10251.

Abstract

A method for pathway selection in a multiple spin-echo pulse sequence applying crusher gradients before and after each pi pulse, to dephase unwanted pathways, is described. This method selects the only pathway that would contribute to the measurable signal if the pi pulses were perfect (1,-1,1,-1, em leader ). Good pathway selection is essential in pulse programming, especially when the CPMG condition is not met. An interactive applet was also developed to deal with these calculations.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Magnetic Resonance Spectroscopy / methods*
  • Mathematics