Effect of the degree of tricuspid regurgitation on cardiac output measurements by thermodilution

Intensive Care Med. 2002 Aug;28(8):1117-21. doi: 10.1007/s00134-002-1352-0. Epub 2002 Jun 15.

Abstract

Objective: To determine the relationship between the degree of tricuspid regurgitation (TR) and accuracy of cardiac output measurement by thermodilution in mechanically ventilated patients. DESIGN AND SETTING. Prospective observational study in a 20-bed general intensive care unit in the university hospital.

Patients: We examined 27 patients (not undergoing cardiac surgery): 8 with no or 1st degree TR, 9 with 2nd degree, and 10 with 3rd degree TR.

Interventions: All patients were measured twice using simultaneously transesophageal echocardiography and pulmonary artery catheter for cardiac output.

Measurements and results: Continuous Doppler measurements were taken in the left ventricular outflow tract at the level of the aortic valve. Cardiac output was calculated by multiplying the velocity-time integral by aortic valve area and heart rate. Simultaneous pulmonary artery catheter measurements were taken averaging the results of the three 10-cc boluses of iced saline. The difference between the methods was 0.5+/-1.1 l/min (mean +/-2 SD) in patients with no or 1st degree TR (r=0.96), 0.8+/-2.0 l/min in those with 2nd degree TR (r=0.92), and 1.9+/-2.3 l/min in those with 3rd degree TR (r=0.69).

Conclusions: A high degree of TR is associated with underestimation of cardiac output measured by thermodilution.

Publication types

  • Comparative Study

MeSH terms

  • Cardiac Output*
  • Catheterization, Swan-Ganz
  • Czech Republic
  • Echocardiography
  • Female
  • Humans
  • Male
  • Prospective Studies
  • Respiration, Artificial
  • Thermodilution / methods*
  • Tricuspid Valve Insufficiency / diagnostic imaging
  • Tricuspid Valve Insufficiency / physiopathology*
  • Tricuspid Valve Insufficiency / therapy
  • Ultrasonography, Doppler