Experimental assessment of precision and accuracy of radiostereometric analysis for the determination of polyethylene wear in a total hip replacement model

J Orthop Res. 2002 Jul;20(4):688-95. doi: 10.1016/S0736-0266(01)00171-1.

Abstract

The purpose of this study was to develop and test a phantom model based on actual total hip replacement (THR) components to simulate the true penetration of the femoral head resulting from polyethylene wear. This model was used to study both the accuracy and the precision of radiostereometric analysis, RSA, in measuring wear. We also used this model to evaluate optimum tantalum bead configuration for this particular cup design when used in a clinical setting. A physical model of a total hip replacement (a phantom) was constructed which could simulate progressive, three-dimensional (3-D) penetration of the femoral head into the polyethylene component of a THR. Using a coordinate measuring machine (CMM) the positioning of the femoral head using the phantom was measured to be accurate to within 7 microm. The accuracy and precision of an RSA analysis system was determined from five repeat examinations of the phantom using various experimental set-ups of the phantom. The accuracy of the radiostereometric analysis, in this optimal experimental set-up studied was 33 microm for the medial direction, 22 microm for the superior direction, 86 microm for the posterior direction and 55 microm for the resultant 3-D vector length. The corresponding precision at the 95% confidence interval of the test results for repositioning the phantom five times, measured 8.4 microm for the medial direction, 5.5 microm for the superior direction, 16.0 microm for the posterior direction, and 13.5 microm for the resultant 3-D vector length. This in vitro model is proposed as a useful tool for developing a standard for the evaluation of radiostereometric and other radiographic methods used to measure in vivo wear.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arthroplasty, Replacement, Hip*
  • Biocompatible Materials / chemistry*
  • Femur Head / diagnostic imaging*
  • Humans
  • Phantoms, Imaging
  • Photogrammetry
  • Polyethylene / chemistry*
  • Radiography

Substances

  • Biocompatible Materials
  • Polyethylene