[Evaluation of the system for controlling the quality of the HBsAg detection in the system of screening laboratories]

Zh Mikrobiol Epidemiol Immunobiol. 2002 May-Jun:(3):52-6.
[Article in Russian]

Abstract

The two-stage control system, ensuring the high quality of serological investigations in the network of screening laboratories in Moscow was developed and introduced into practice. At the first stage the entry control of the quality of the test system for the detection of HbsAg, coming to the screening laboratories, is made in the reference laboratory with the use of specially developed "representative" panels, as well as the test systems for comparison ("reference" test systems). Then "minipanels" are formed from specimens included into the "representative" panels, which are used for the evaluation of the quality of laboratory investigations made in the screening laboratories. The high quality of such system for controlling the quality of the detection of HBsAg in the screening laboratories of Moscow is shown.