Experimental investigations of image quality in X-ray mammography with conventional screen film system (SFS), digital phosphor storage plate in/without magnification technique (CR) and digital CCD-technique (CCD)

Rontgenpraxis. 2001;54(4):123-6.

Abstract

Comparison of image quality in X-ray mammography between conventional film screen film system (SFS), digital phosphor storage plate in and without magnification technique (CR) and digital CCD-technique (CCD). Radiograms of an RMI-mammography phantom were acquired using a conventional screen film system, three digital storage plate systems and two digital systems in CCD-technique. Additionally, the radiograms of one digital phosphor storage plate system were post-processed regarding contrast and included in the comparison. The detectability of details was best with the digital mammography in CCD-technique. After confirming these promising results in clinical studies, digital mammography should be able to replace conventional screen film technique.

Publication types

  • Comparative Study

MeSH terms

  • Breast Neoplasms / diagnostic imaging*
  • Calcinosis / diagnostic imaging*
  • Female
  • Humans
  • Mammography / instrumentation*
  • Phantoms, Imaging
  • Radiographic Image Enhancement*
  • Radiographic Magnification*
  • Sensitivity and Specificity
  • X-Ray Intensifying Screens*