Using thermogravimetry for weight loss monitoring of permeation tubes used for generation of trace concentration gas standards

Anal Chem. 2002 Jan 1;74(1):305-7. doi: 10.1021/ac0108921.

Abstract

Permeation tubes are convenient analyte sources for generating standard gas mixtures (containing, in particular, volatile organic compounds) used in the calibration of analytical instruments. For small permeation rates, corresponding to trace levels of analytes, the calibration of permeation tubes is time-consuming. The use of thermogravimetry as a means to measure rapidly the weight loss at constant temperature was investigated. An attempt to apply this technique to calibrating a permeation tube filled with benzene is described. In the 20 ng/min range of permeation rate, day-to-day variations of <5% were observed. The continuous weight loss monitoring of permeation tubes by thermogravimetry allows their rapid characterization.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air Pollutants / analysis*
  • Air Pollutants / standards
  • Environmental Monitoring / instrumentation*
  • Environmental Monitoring / standards
  • Gases / standards
  • Organic Chemicals / analysis
  • Reference Standards

Substances

  • Air Pollutants
  • Gases
  • Organic Chemicals