Novel femoral artery puncture closure device in patients undergoing interventional and diagnostic cardiac procedures

J Invasive Cardiol. 2002 Jan;14(1):9-12.

Abstract

Background: The wide range of closure devices for arterial access sites still leaves room for improvement.

Methods and results: We report our initial experience with a novel, cost-effective, suture-mediated device (6 French X-PRESS device, X-SITE Medical, Blue Bell, Pennsylvania) for closure of the arterial access site after both diagnostic and therapeutic interventions. A total of 51 consecutive patients are reported. The closure was performed immediately after the procedure. No patient had more than 5,000 U heparin administered during the procedure. The closure device was used in 36 patients (73%) after diagnostic angiography and in 15 patients (27%) after therapeutic intervention. The device could be successfully deployed in 48 patients (94%). In the remaining 3 cases (6%), hemostasis was achieved with standard manual compression without any further event. Immediate total hemostasis was achieved with the device in 44 of 48 patients (92%). Four devices (8%) could not be deployed correctly due to technical problems and the patients needed additional manual compression. The average time to achieve complete hemostasis was 5.5 +/- 3.5 minutes (range, 3-26 minutes). The patients were ambulated after a period of 1.3 +/- 0.4 hours of bedrest (range, 0.8-3.5 hours). Time to possible discharge was 4.2 +/- 3.4 hours (range, 0.9-12.5 hours). One patient experienced a minor complication (hematoma < 6 cm) and was treated with manual compression without clinical sequelae. Telephone follow-up was carried out 2 weeks after deployment and revealed no sequelae.

Conclusion: The 6 French X-PRESS device is safe and effective in providing rapid hemostasis following interventional or diagnostic catheterization procedures. Use of the device is associated with a low rate of complications and facilitates quick hemostasis, mobilization and discharge.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Cardiac Surgical Procedures / instrumentation*
  • Diagnostic Techniques, Cardiovascular*
  • Early Ambulation
  • Equipment Design / instrumentation
  • Equipment Safety / instrumentation
  • Female
  • Femoral Artery / surgery*
  • Follow-Up Studies
  • Hemostasis, Surgical / instrumentation
  • Humans
  • Male
  • Prospective Studies
  • Punctures / instrumentation*
  • Suture Techniques / instrumentation
  • Treatment Outcome