High-throughput proteomics using high-efficiency multiple-capillary liquid chromatography with on-line high-performance ESI FTICR mass spectrometry

Anal Chem. 2001 Jul 1;73(13):3011-21. doi: 10.1021/ac001393n.

Abstract

We report on the design and application of a high-efficiency multiple-capillary liquid chromatography (LC) system for high-throughput proteome analysis. The multiple-capillary LC system using commercial LC pumps was operated at a pressure of 10,000 psi to deliver mobile phases through a novel passive feedback valve arrangement that permitted mobile-phase flow path switching and efficient sample introduction. The multiple-capillary LC system uses several serially connected dual-capillary column devices. The dual-capillary column approach eliminates the time delays for column regeneration (or equilibration) since one capillary column was used for a separation while the other was being washed. Several serially connected dual-capillary columns and electrospray ionization (ESI) sources were operated independently and can be used either for "backup" operation or for parallel operation with other mass spectrometers. This high-efficiency multiple-capillary LC system utilizes switching valves for all operations, enabling automated operation. The separation efficiency of the dual-capillary column arrangement, optimal capillary dimensions (column length and packed particle size), capillary regeneration conditions, and mobile-phase compositions and their compatibility with electrospray ionization were investigated. A high magnetic field (11.4 T) Fourier transform ion cyclotron resonance (FTICR) mass spectrometer was coupled on-line with this high-efficiency multiple-capillary LC system using an ESI interface. The capillary LC provided a peak capacity of approximately 650, and the 2-D capillary LC-FTICR analysis provided a combined resolving power of > 6 x 10(7) components. For yeast cytosolic tryptic digests > 100,000 polypeptides were detected, and approximately 1,000 proteins could be characterized from a single capillary LC-FTICR analysis using the high mass measurement accuracy (approximately 1 ppm) of FTICR, and likely more if LC retention time information were also exploited for peptide identification.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Chromatography, Liquid / methods*
  • Fungal Proteins / chemistry
  • Peptide Mapping
  • Proteome*
  • Reproducibility of Results
  • Saccharomyces cerevisiae / chemistry
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Spectrophotometry, Ultraviolet
  • Spectroscopy, Fourier Transform Infrared
  • Trypsin / chemistry

Substances

  • Fungal Proteins
  • Proteome
  • Trypsin