Trace level determination of phenols as pentafluorobenzyl derivatives by gas chromatography-negative-ion chemical ionization mass spectrometry

Analyst. 2001 Jun;126(6):835-9. doi: 10.1039/b101194o.

Abstract

A method for the trace level determination of 11 phenols as pentafluorobenzyl (PFB) derivatives by gas chromatography-mass spectrometry (GC-MS) with negative-ion chemical ionization (NICI) is described. First, the conditions for the PFB derivatisation of phenols were optimized and were found to be reaction temperature 80 degrees C and reaction time 5 h. Second, the detection limits using selected ion monitoring (SIM) were compared between trimethylsilylated (TMS) derivatives in the electron ionization (EI) mode and PFB derivatives in the NICI mode. The responses for the PFB derivatives in the NICI mode were 3.3-61 times higher than those of the TMS derivatives in the EI mode. The instrumental detection limits using NICI-SIM ranged from 2.6 to 290 fg. This method was applied to the analysis of phenols in river water using solid-phase extraction. The recoveries of the phenols from a river water sample spiked with standards at 100 ng l-1 with 2-chlorophenol, 4-chloro-3-methylphenol and pentachlorophenol and at 1000 ng l-1 with phenol, 2,4-dimethylphenol, 2,4-dichlorophenol, 2-nitrophenol, 2,4,6-trichlorophenol and 4-nitrophenol were 81.2-106.3% (RSD 5.1-8.0%), except for 2-methyl-4,6-dinitrophenol and 2,4-dinitrophenol, for which the recoveries were 5.8 and 4.2%, respectively, because water contained in the acetone eluate interfered with the derivatisation of these compounds with two electrophilic nitro groups.

MeSH terms

  • Gas Chromatography-Mass Spectrometry
  • Phenols / analysis*
  • Water Pollutants, Chemical / analysis*

Substances

  • Phenols
  • Water Pollutants, Chemical