Magnetic resonance imaging system for three-dimensional conformal radiotherapy and its impact on gross tumor volume delineation of central nervous system tumors

Int J Radiat Oncol Biol Phys. 2001 Jul 1;50(3):821-7. doi: 10.1016/s0360-3016(01)01598-x.

Abstract

Purpose: We developed an MRI system for three-dimensional planning in radiotherapy. Its contribution on gross tumor volume (GTV) delineation of central nervous system (CNS) diseases was evaluated.

Methods and materials: The MRI system, with corrected distortion, was registered on computed tomography (CT) by means of fiducial/anatomic landmarks. In 41 consecutive patients with various CNS diseases, GTVs determined by MRI/CT registration (MR/CT-GTV) and CT alone (CT-GTV) were compared. Hard copies of diagnostic MRI were shown to doctors when CT-GTV was determined to simulate a conventional planning situation. Multi-observer volumetric analysis was conducted, assessing interobserver deviations among four radiation oncologists and intermethodological deviations between MR/CT-GTV and CT-GTV.

Results: Overall, the mean of geometric distortion was significantly reduced from 1.08 mm to 0.3 mm by distortion correction (p < 0.0001). The contribution of the correction was apparent at >12.0 cm radius from the center of the magnetic field. Interobserver deviation was significantly reduced by MR/CT registration (p = 0.005). The improvement was significant for acoustic neurinoma (p = 0.038), astrocytomas (p = 0.043), and lesions at the cerebellum/brainstem (p = 0.008). The regression coefficient between MR/CT-GTV and CT-GTV was <0.9 for cerebellum/brainstem lesions, suggesting that MRI/CT-GTV was smaller than CT-GTV.

Conclusions: This system is feasible for three-dimensional planning and was shown to reduce interobserver deviations in GTV delineation for CNS diseases.

Publication types

  • Clinical Trial

MeSH terms

  • Brain Neoplasms / pathology
  • Brain Neoplasms / radiotherapy*
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Conformal / methods
  • Tomography, X-Ray Computed