Selective trace level analysis of phenolic compounds in water by flow injection analysis--membrane introduction mass spectrometry

Environ Sci Technol. 2001 May 15;35(10):2084-8. doi: 10.1021/es001814i.

Abstract

Flow injection analysis coupled with membrane introduction mass spectrometry (FIA-MIMS) with on-line derivatization is shown to allow fast, accurate, nearly interference-free, and sensitive (low microgram/L) quantitation of phenolic compounds in water. On-line FIA derivatization of the phenolic compounds is performed by acetic anhydride acetylation in a K2CO3-buffered alkaline medium. The phenol acetates so formed efficiently permeate a silicone membrane and are directly transferred to the mass spectrometer, in which they are analyzed with selectivity and high sensitivity via selected ion monitoring. FIA-MIMS analysis was performed for aqueous solutions of phenol, 2-methylphenol, 4-chlorophenol, 4-chloro-3-methylphenol, 2,4-dichlorophenol, and 2,4,6-trichlorophenol, and detection limits in the 0.5-20 micrograms/L (ppb) range were observed for an analytical frequency of six samples/h. FIA-MIMS for phenolic compound analysis is considerably less time-consuming and labor intensive than most chromatographic methods based on liquid-liquid extraction and preconcentration procedures and is therefore applicable for on-line and in-situ monitoring of phenols in wastewaters and in the environment. FIA-MIMS employing acetic anhydride derivatization is also virtually free of interferences since it combines chemical, membrane, and enhanced MS selectivity; hence quantitation of phenolic compounds can be performed in the presence of congeners.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acetates / chemistry
  • Environmental Monitoring / methods*
  • Mass Spectrometry / methods*
  • Phenols / analysis*
  • Sensitivity and Specificity
  • Water Pollutants, Chemical / analysis*

Substances

  • Acetates
  • Phenols
  • Water Pollutants, Chemical