Removal of the retained fragment of broken solid nails by the intra-medullary route

Injury. 2001 Jun;32(5):407-10. doi: 10.1016/s0020-1383(00)00236-9.

Abstract

The development of solid nails has not eliminated the risk of occasional nail failure. It has been suggested that subsequent intra-medullary removal of solid nail fragment may be virtually impossible and thus major surgery is required for their revision. We report two cases of hardware failure of a solid intra-medullary nail (Synthes solid femoral and tibial nail) and describe their successful intra-medullary removal using the Synthes extraction kit.

Publication types

  • Case Reports

MeSH terms

  • Adult
  • Bone Nails*
  • Device Removal / methods*
  • Femoral Fractures / surgery*
  • Fracture Fixation, Intramedullary / instrumentation*
  • Humans
  • Male
  • Prosthesis Failure
  • Reoperation
  • Tibial Fractures / surgery*