Evaluation of in vivo dose measurements for patients undergoing electron boost treatments

Radiother Oncol. 2001 Jun;59(3):293-6. doi: 10.1016/s0167-8140(01)00323-1.

Abstract

This study evaluated p-type silicon diodes for use in in vivo dosimetry in clinical electron beams. A calibrated p-type silicon diode detector was used to measure the dose received by the patient in the centre of the field. Readings were corrected for energy, temperature and stand-off of the electron applicator from the patient surface. The mean difference between measured and prescribed dose was 1.04% (95% CI 0.72 to 1.36 %).