Monitoring motility, spreading, and mortality of adherent insect cells using an impedance sensor

Anal Chem. 2001 Apr 15;73(8):1844-8. doi: 10.1021/ac0011585.

Abstract

An emerging sensor technology referred to as electric cell-substrate impedance sensing (ECIS) has been extended for monitoring the behavior of insect cells including attachment, motility, and mortality. In ECIS, adherent cells were cultured on an array of eight small gold electrodes deposited on the bottom of tissue culture wells and immersed in a culture medium. Upon the attachment and spreading of cells on the gold electrode, the impedance increased because the cells acted as insulating particles to restrict the current flow. Experimental data revealed that insect cells interacted differently with various proteins used to precoat the gold electrode with concanavalin A as the best promoter to accelerate the rate of cell attachment. After the cells were fully spread, the measured impedance continued to fluctuate to reflect the constant motion and metabolic activity of the cells. As the cell behavior was sensitive to external chemicals, the applicability of ECIS for inhibition assays was demonstrated with HgCl2, trinitrotoluene, trinitrobenzene (TNB), and 2-amino-4,6-dinitrotoluene as model systems. Unlike conventional assays, the quantitative data obtained in this study are taken in real time and in a continuous fashion to depict cell motility and mortality.

MeSH terms

  • Animals
  • Biosensing Techniques*
  • Cell Adhesion
  • Cell Death
  • Cell Movement
  • Cells, Cultured
  • Electric Impedance
  • Spodoptera / cytology*