Integral method for numerical simulation of MRI artifacts induced by metallic implants

Magn Reson Med. 2001 Apr;45(4):724-7. doi: 10.1002/mrm.1098.

Abstract

Numerical simulation is a valuable tool for the study of magnetic susceptibility artifacts from metallic implants. A major difficulty in the simulation lies in the computation of the magnetic field induced by the metallic implant. A new method has been designed and implemented to compute the magnetic field induced by metallic objects of arbitrary shape. The magnetic field is expressed pointwise in terms of a surface integral. Efficient quadrature schemes are proposed to evaluate this integral. Finally, the method is linked to an artifact reconstruction model to simulate the images. Magn Reson Med 45:724-727, 2001.

MeSH terms

  • Artifacts
  • Magnetic Resonance Imaging / methods*
  • Magnetics
  • Mathematics
  • Metals*
  • Models, Theoretical
  • Prostheses and Implants*

Substances

  • Metals