Interchangeability of gas detection tubes and hand pumps

AIHAJ. 2001 Jan-Feb;62(1):65-9. doi: 10.1080/15298660108984611.

Abstract

Users of gas detection tubes occasionally seek the convenience of using a single hand pump with different brands of tubes, to avoid the need to carry more than one pump. Several professional organizations recommend against such interchange. However, these recommendations appear to be based on a single study of pump designs that mostly are no longer in use. The present study was undertaken to determine if current hand pumps are interchangeable. Both piston-type and bellows-type hand pumps were evaluated by comparing pump flow profiles and test gas measurements with a variety of tubes. The results demonstrate that three piston hand pumps in common use (Sensidyne/Gastec GV/100, RAE Systems LP-1200, and Matheson-Kitagawa 8104-400A) are fully interchangeable. Two bellows pumps (Draeger Accuro and MSA Kwik-Draw) also are interchangeable with each other. Mixing of bellows and piston systems is often possible, but there are enough exceptions to conclude that such practice should be discouraged because it can give inaccurate readings. It is recommended that technical standards be adopted, such as total volume and an initial pump vacuum or a pump flow curve, to assess hand pump interchangeability. When two manufacturers' pumps meet the same standard and routine leak tests are conducted, interchangeability is scientifically valid and poses no risk to the end user while offering greater convenience.

Publication types

  • Comparative Study

MeSH terms

  • Air Pollutants / analysis*
  • Colorimetry / instrumentation
  • Equipment Design*
  • Gases / analysis*
  • United States
  • Vacuum Curettage / standards*

Substances

  • Air Pollutants
  • Gases