Evaluation of linear and nonlinear tomosynthetic reconstruction methods in digital mammography

Acad Radiol. 2001 Mar;8(3):219-24. doi: 10.1016/S1076-6332(03)80530-5.

Abstract

Rationale and objectives: The purpose of this study was to comparatively evaluate digital planar mammography and both linear and nonlinear tomosynthetic reconstruction methods.

Materials and methods: A "disk" (ie, target) identification study was conducted to compare planar and reconstruction methods. Projective data using a composite phantom with circular disks were acquired in both planar and tomographic modes by using a full-field, digital mammographic system. Two-dimensional projections were reconstructed with both linear (ie, backprojection) and nonlinear (ie, maximization and minimization) tuned-aperture computed tomographic (TACT) methods to produce three-dimensional data sets. Four board-certified radiologists and one 4th-year radiology resident participated as observers. All images were compared by these observers in terms of the number of disks identified.

Results: Significant differences (P < .05, Bonferroni adjusted) were observed between all reconstruction and planar methods. No significant difference, however, was observed between the planar methods, and only a marginally significant difference (P < .054, Bonferroni adjusted) was observed between TACT-backprojection and TACT-minimization.

Conclusion: A combination of linear and nonlinear reconstruction schemes may have potential implications in terms of enhancing image visualization to provide radiologists with valuable diagnostic information.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Breast Neoplasms / diagnostic imaging
  • Female
  • Humans
  • Mammography / methods*
  • Phantoms, Imaging
  • Radiographic Image Enhancement*
  • Sensitivity and Specificity