On-line particle concentrator with upstream ultrafiltration in continuous SPLITT fractionation

Anal Chem. 2001 Feb 1;73(3):693-7. doi: 10.1021/ac0009550.

Abstract

An upstream ultrafiltration (UU) method is employed for the on-line concentration of collected particle fraction and for the convenient regulation of flow rates in split-flow thin (SPLITT) fractionation. Concentration of the collected particle fraction is necessary to fractionate particles by smaller cutoff diameters in SPLITT fractionation. By introducing a simple device utilizing upstream ultrafiltration with tangential flow, particle solution collected at each SP