Design optimization of a breast imaging system based on silicon microstrip detectors

Phys Med Biol. 2000 Jul;45(7):2029-41. doi: 10.1088/0031-9155/45/7/323.

Abstract

A mammographic imaging set-up using silicon microstrip detectors in edge-on geometry was simulated using the GEANT package. Deposited energy in tissue of various thicknesses was evaluated and shown to agree to within 10% with reference calculations. Optimal energies as well as spectra for mammography with silicon detectors were determined by maximizing the figure of merit of a realistic imaging set-up. The scattered to primary radiation ratio was studied for various detector geometries. It was found that fan-shaped detectors are needed to maintain the image quality for divergent photon beams.

MeSH terms

  • Computer Simulation
  • Humans
  • Mammography / instrumentation*
  • Mammography / methods*
  • Models, Statistical
  • Photons
  • Silicon
  • Software
  • X-Rays

Substances

  • Silicon