Dynamic system for the calibration of semi-volatile carbonyl compounds in air

J Chromatogr A. 2000 Jun 23;883(1-2):171-83. doi: 10.1016/s0021-9673(00)00379-4.

Abstract

A dynamic calibration system designed for generating accurate and precise concentrations of semi-volatile carbonyls in the ppb (v/v) (ppbv)-ppt (v/v) (pptv) range has been developed and tested. Alkanals from C6 to C9 were used as test compounds. Diffusion rates of their vapours from capillary tubes were determined theoretically and with two methods whose accuracy was independently evaluated with liquid standard solutions. Methods selected for testing the calibration system were those commonly used for the selective quantification of carbonyls in air. One is based on the well-known reaction with 2,4-dinitrophenylhydrazine (DNPH), followed by the analysis of formed hydrazones by HPLC-UV. The other is based on the retention of carbonyls on graphitic carbon adsorbents, followed by MS detection of the sample injected into a capillary column by thermal desorption. A good consistency was found between the values of the diffusion rates determined by the two methods. The scatter between the two methods was in the range of +/-10% when diffusion rates of ca. 1 ng/s were attained. Experimental values closely approached those calculated by applying the equation describing the diffusion of a vapour in equilibrium with the liquid through a capillary tube.

MeSH terms

  • Air / analysis*
  • Aldehydes / analysis*
  • Calibration
  • Chromatography, High Pressure Liquid
  • Gas Chromatography-Mass Spectrometry
  • Spectrophotometry, Ultraviolet