Purpose: To evaluate the feasibility and efficacy of a newly designed stent-graft placed across the renal arteries for exclusion of abdominal aortic aneurysms (AAAs) with short or tortuous proximal necks.

Methods: Among a group of AAA patients treated with endovascular grafting, 5 had tortuous proximal necks and 13 had necks <20 mm (mean 13 mm). In these 18 cases, a 2- to 3-cm uncovered segment of the stent-graft was placed transrenally using a catheter inserted into the renal artery as a guide for graft margin positioning. A newly designed stent-graft was constructed from a custom-made spiral Z-stent covered with a thin-walled Dacron material; the endografts were deployed through 16-F (aortoaortic model) or 18-F sheaths (bifurcated devices). Renal function was assessed by preoperative and postoperative measurement of urea nitrogen and creatinine. Aneurysm exclusion and renal artery patency were evaluated during follow-up using spiral computed tomography and angiography.

Results: The stent-grafts were correctly placed at the intended site in all 18 patients. Renal function was not affected except transiently in 1 patient who developed bilateral renal artery stenoses 24 hours after the procedure; Palmaz stents were deployed in each renal artery to reestablish satisfactory blood flow. Of the 33 renal arteries crossed by the bare stent-graft segment, all were patent over a mean 14-month follow-up (range 7-24), including the patient with Palmaz stents implanted for postprocedural renal stenosis. Complete aneurysm exclusion was maintained in 15 (83%) of 18 patients; proximal leaks persisted in 3 patients, including 2 with severely angled proximal necks.

Conclusions: Transrenal placement of the uncovered leading edge of custom-made spiral Z-stent-based endografts appears feasible and clinically effective in the treatment of AAAs with short or tortuous proximal necks.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Angiography, Digital Subtraction
  • Aortic Aneurysm, Abdominal / diagnostic imaging
  • Aortic Aneurysm, Abdominal / surgery*
  • Blood Vessel Prosthesis Implantation / instrumentation*
  • Blood Vessel Prosthesis*
  • Coated Materials, Biocompatible*
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Polyethylene Terephthalates
  • Prosthesis Design
  • Renal Artery
  • Retrospective Studies
  • Stainless Steel
  • Stents*
  • Tomography, X-Ray Computed
  • Ultrasonography, Doppler

Substances

  • Coated Materials, Biocompatible
  • Polyethylene Terephthalates
  • Stainless Steel