Reconstructing MR images from undersampled data: data-weighting considerations

Magn Reson Med. 2000 Jun;43(6):867-75. doi: 10.1002/1522-2594(200006)43:6<867::aid-mrm13>3.0.co;2-2.

Abstract

Data which are sampled more densely than the Nyquist limit in k-space are weighted prior to reconstruction by the inverse of the local sampling density. This work considers the effects of weighting data that are sampled less densely than the Nyquist limit. It specifically analyzes azimuthally undersampled projection reconstruction, variable density spirals, and variable density phase encoding. Effects on resolution, aliasing, and SNR are given. Higher resolution is obtained by weighting undersampled data according to the inverse of sampling density, while better SNR and less aliasing artifact are obtained by weighting undersampled data uniformly. Magn Reson Med 43:867-875, 2000.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / methods*
  • Models, Theoretical*
  • Sensitivity and Specificity