American College of Cardiology/European Society of Cardiolgoy International Study of Angiographic Data Compression Phase II: the effects of varying JPEG data compression levels on the quantitative assessment of the degree of stenosis in digital coronary angiography. Joint Photographic Experts Group

J Am Coll Cardiol. 2000 Apr;35(5):1380-7. doi: 10.1016/s0735-1097(99)00611-7.

Abstract

Objectives: This report describes whether lossy Joint Photographic Experts Group (UPEG) image compression/decompression has an effect on the quantitative assessment of vessel sizes by state-of-the-art quantitative coronary arteriography (QCA).

Background: The Digital Imaging and Communications in Medicine (DICOM) digital exchange standard for angiocardiography prescribes that images must be stored loss free, thereby limiting JPEG compression to a maximum ratio of 2:1. For practical purposes it would be desirable to increase the compression ratio (CR), which would lead to lossy image compression.

Methods: A series of 48 obstructed coronary segments were compressed/decompressed at CR 1:1 (uncompressed), 6:1, 10:1 and 16:1 and analyzed blindly and in random order using the QCA-CMS analytical software. Similar catheter and vessel start- and end-points were used within each image quartet, respectively. All measurements were repeated after several weeks using newly selected start- and end-points. Three different sub-analyses were carried out: the intra-observer, fixed inter-compression and variable inter-compression analyses, with increasing potential error sources, respectively.

Results: The intra-observer analysis showed significant systematic and random errors in the calibration factor at JPEG CR 10:1. The fixed inter-compression analysis demonstrated systematic errors in the calibration factor and recalculated vessel parameter results at CR 16:1 and for the random errors at CR 10:1 and 16:1. The variable inter-compression analysis presented systematic and random errors in the calibration factor and recalculated parameter results at CR 10:1 and 16:1. Any negative effect at CR 6:1 was found only for the calibration factor of the variable inter-compression analysis, which did not show up in the final vessel measurements.

Conclusions: Compression ratios of 10:1 and 16:1 affected the QCA results negatively and therefore should not be used in clinical research studies.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aged
  • Analysis of Variance
  • Bias
  • Calibration
  • Cardiology
  • Coronary Angiography / methods*
  • Coronary Angiography / standards*
  • Coronary Disease / classification*
  • Coronary Disease / diagnostic imaging*
  • Europe
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Image Processing, Computer-Assisted / standards*
  • Middle Aged
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Severity of Illness Index*
  • Single-Blind Method
  • Societies, Medical
  • United States