Development of membrane extraction with a sorbent interface-micro gas chromatography system for field analysis

J Chromatogr A. 2000 Mar 17;873(1):13-27. doi: 10.1016/s0021-9673(99)01318-7.

Abstract

The commercially available portable gas chromatographs have a rather limited scope of applications, typically allowing analysis of gaseous samples only, and having relatively poor sensitivity. Combination of those instruments with modern sampling/sample preparation techniques can remedy these problems. A Chrompack micro-GC system equipped with a thermal conductivity detector has been coupled to membrane extraction with a sorbent interface (MESI). The sorbent trap has replaced the GC injector. The design of the trap was also modified in order to enhance the preconcentration of analytes. The use of a thin flat sheet membrane reduces the response time, and decreases the memory effect of the system. Rapid separation times were achieved, and the sensitivity was significantly improved. MESI enables semi-continuous monitoring of both gaseous and aqueous samples, owing to the selectivity of the membrane material. The system does not use moving parts, therefore being reliable. The sensitivity of the micro-GC system was increased by a factor of more than 100 by the addition of the MESI system, even with a preconcentration time as short as 1 min. Chloroform, having a concentration lower than 1 ppb, was detected in tap water. A cup system was used to allow headspace sampling of volatile organic compounds from aqueous matrices, keeping the membrane away from interfering species that could be present in water, and improving the mass transfer. A linear calibration line was obtained, and the estimated limit of detection was 60 ppt. This represents a great improvement for the sensitivity of the micro-GC system.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, Gas / methods*
  • Membranes, Artificial
  • Sensitivity and Specificity

Substances

  • Membranes, Artificial