Screen-printed amperometric biosensors for the rapid measurement of L- and D-amino acids

Analyst. 1999 Jun;124(6):865-70. doi: 10.1039/a901404g.

Abstract

Screen-printed three-electrode amperometric sensors incorporating L- and/or D-amino acid oxidase for the general purpose measurement of L- or D-amino acids is described. The working electrode incorporates rhodinized carbon, to facilitate hydrogen peroxide oxidation at a decreased operating potential, and immobilized enzyme. The devices responded to all 20 common L-amino acids and all of the D-amino acids examined, the exceptions being L- and D-proline. Linear response profiles were observed for L-leucine, L-glycine and L-phenylalanine with limits of detection of 0.47, 0.15 and 0.20 mM respectively. The devices were reproducible and exhibited stability over a 56 d test period. The biosensor compares favourably with a standard photometric amino acid test and was used to monitor milk ageing effects. The assay is cheap, simple to perform and rapid, requiring only buffer-electrolyte and a small sample volume.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Amino Acids / analysis*
  • Animals
  • Biomarkers / analysis
  • Biosensing Techniques*
  • Cattle
  • Food Microbiology*
  • Isomerism
  • Milk / microbiology*

Substances

  • Amino Acids
  • Biomarkers