Two reproducible and sensitive liquid chromatographic methods to quantify atenolol and propranolol in human plasma and determination of their associated analytical error functions

J Chromatogr B Biomed Sci Appl. 2000 Feb 11;738(2):225-31. doi: 10.1016/s0378-4347(99)00522-8.

Abstract

Two liquid chromatography (LC) methods with fluorimetric detection have been developed to measure atenolol and propranolol in human plasma. The same 5 microm Nucleosil RP-18 column, extraction procedure and mobile phase (containing acetonitrile, water, triethylamine and phosphoric acid, pH 3) were used. The linearity ranges were 25-800 ng/ml for atenolol and 3.13-100 ng/ml for propranolol. The coefficients of variation for validation assays were lower than 15% at the concentration assayed. The functions of the analytical error were linear: SD (ng/ml)=7.698+0.037C for atenolol and SD (ng/ml)=0.126+0.036C for propranolol.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adrenergic beta-Antagonists / blood*
  • Atenolol / blood*
  • Chromatography, Liquid / methods*
  • Humans
  • Propranolol / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Fluorescence

Substances

  • Adrenergic beta-Antagonists
  • Atenolol
  • Propranolol