Analytical and clinical performance of two cardiac troponin I immunoassays

Clin Chem Lab Med. 1999 Sep;37(9):889-97. doi: 10.1515/CCLM.1999.132.

Abstract

Cardiac troponin I assays for Axsym (Abbott Diagnostics, Abbott Park, IL, USA) and Immuno 1 (Bayer Corporation, Tarrytown, NY, USA) analysers were evaluated. Heparin plasma or serum could be used for both assays. Samples were stable for 24 h at ambient temperature, 3 days at 4-8 degrees C and 3 months at -20 degrees C. After 10 months' storage at -80 degrees C, the recoveries were well above 100% by both assays. Total coefficients of variation for Axsym assay were 9.0%, 5.8% and 5.3% at concentrations of 2.6 microg/l, 9.83 microg/l and 34.3 microg/l respectively; for Immuno 1 these were 4.4 %, 1.6% and 1.8% at 2.3 microg/l, 6.27 microg/l and 44.35 microg/l respectively. It was > or =20% at concentration of < or =0.5 microg/l for Axysm assay and < or =0.15 microg/l for Immuno 1 assay. Recoveries were < or =90% at < or =0.22 microg/l on Axsym and at < or =1.47 microg/l on Immuno 1. Neither method showed significant interference with haemoglobin, bilirubin, triglycerides or rheumatoid factor. Correlation between the two methods was excellent (r = 0.997, Y (Axsym) = 4.2X (Immuno 1) +3.2). The highest concentrations detected in 50 healthy subjects were 0.3 microg/l and 0.1 microg/l by Axsym and Immuno 1 methods, respectively. Twelve out of 43 renal failure patients had troponin I 0.13-0.9 microg/l using Axsym method and 4 had levels of 0.07-0.13 microg/l using Immuno 1. In muscle trauma patients, troponin I was undetectable.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Biomarkers
  • Female
  • Humans
  • Immunoassay / methods*
  • Kidney Failure, Chronic / metabolism
  • Male
  • Middle Aged
  • Muscles / injuries
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Troponin I / analysis*

Substances

  • Biomarkers
  • Troponin I