Comparison and evaluation of rigid, affine, and nonrigid registration of breast MR images

J Comput Assist Tomogr. 1999 Sep-Oct;23(5):800-5. doi: 10.1097/00004728-199909000-00031.

Abstract

Purpose: A new nonrigid registration method, designed to reduce the effect of movement artifact in subtraction images from breast MR, is compared with existing rigid and affine registration methods.

Method: Nonrigid registration was compared with rigid and affine registration methods and unregistered images using 54 gadolinium-enhanced 3D breast MR data sets. Twenty-seven data sets had been previously reported normal, and 27 contained a histologically proven carcinoma. The comparison was based on visual assessment and ranking by two radiologists.

Results: When analyzed by two radiologists independently, all three registration methods gave better-quality subtraction images than unregistered images (p < 0.01), but nonrigid registration gave significantly better results than the rigid and affine registration methods (p < 0.01). There was no significant difference between rigid and affine registration methods.

Conclusion: Nonrigid registration significantly reduces the effects of movement artifact in subtracted contrast-enhanced breast MRI. This may enable better visualization of small tumors and those within a glandular breast.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Breast / pathology*
  • Contrast Media
  • Female
  • Gadolinium DTPA
  • Humans
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / statistics & numerical data
  • Middle Aged
  • Observer Variation
  • Statistics, Nonparametric

Substances

  • Contrast Media
  • Gadolinium DTPA