CT-Based interstitial HDR brachytherapy

Strahlenther Onkol. 1999 Sep;175(9):419-27. doi: 10.1007/s000660050031.

Abstract

Purpose: Development, application and evaluation of a CT-guided implantation technique and a fully CT-based treatment planning procedure for brachytherapy.

Methods and materials: A brachytherapy procedure based on CT-guided implantation technique and CT-based treatment planning has been developed and clinical evaluated. For this purpose a software system (PROMETHEUS) for the 3D reconstruction of brachytherapy catheters and patient anatomy using only CT scans has been developed. An interface for the Nucletron PLATO BPS treatment planning system for optimization and calculation of dose distribution has been devised. The planning target volume(s) are defined as sets of points using contouring tools and are used for optimization of the 3D dose distribution. Dose-volume histogram based analysis of the dose distribution (COIN analysis) enables a clinically realistic evaluation of the brachytherapy application to be made. The CT-guided implantation of catheters and the CT-based treatment planning procedure has been performed for interstitial brachytherapy and for different tumor sites in 197 patients between 1996 and 1997.

Results: The accuracy of the CT reconstruction was tested using first a quality assurance phantom and second, a simulated interstitial implant of 12 needles. These were compared with the results of reconstruction using radiographs. Both methods gave comparable results with regard to accuracy, but the CT based reconstruction was faster. Clinical feasibility was proved in pre-irradiated recurrences of brain tumors, in pretreated recurrences or metastatic disease, and in breast carcinomas. The tumor volumes treated were in the range 5.1 to 2,741 cm3. Analysis of implant quality showed a slightly significant lower COIN value for the bone implants, but no differences with respect to the planning target volume.

Conclusions: The Offenbach system, incorporating the PROMETHEUS software for interstitial HDR brachytherapy has proved to be extremely valuable in routine clinical practice for many tumor sites. Our CT-guided implantation technique together with a fully CT-based planning system has enabled conformal brachytherapy treatment to become routine.

MeSH terms

  • Brachytherapy / instrumentation*
  • Humans
  • Image Processing, Computer-Assisted / instrumentation*
  • Neoplasms / radiotherapy*
  • Phantoms, Imaging
  • Quality Assurance, Health Care
  • Radiotherapy Planning, Computer-Assisted / instrumentation*
  • Radiotherapy, High-Energy
  • Software
  • Tomography, X-Ray Computed / instrumentation*
  • Treatment Outcome