Assessment of the accuracy of an intermittent-scanning continuous glucose monitoring device in patients with type 2 diabetes mellitus undergoing hemodialysis (AIDT2H) study

Ther Apher Dial. 2021 Oct;25(5):586-594. doi: 10.1111/1744-9987.13618. Epub 2021 Feb 2.

Abstract

FreeStyle Libre has been approved for use in patients undergoing hemodialysis (HD) in Japan, unlike Europe and the United States; however, evidence regarding its accuracy in such patients is sparse. Forty-one participants with type 2 diabetes undergoing HD were recruited. The overall mean absolute relative difference and mean absolute difference were 23.4% and 33.9 mg/dL, respectively. Sensor glucose levels and capillary glucose levels were significantly correlated (r = 0.858, P < .01), although the sensor glucose levels were significantly lower than the capillary glucose levels. The accuracy of FreeStyle Libre in patients undergoing HD became deteriorated with the days of usage. The percentage of sensor results in Zones A and B in the consensus error grid analysis and in the Clarke error grid analysis were 99.7% and 99.0%, respectively. Its insufficient accuracy necessitates adjunct usage of FreeStyle Libre with self-monitoring of blood glucose in patients undergoing HD.

Keywords: FreeStyle Libre; accuracy; continuous glucose monitoring; diabetic nephropathy; hemodialysis; type 2 diabetes.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / methods*
  • Diabetes Mellitus, Type 2 / blood*
  • Female
  • Humans
  • Male
  • Prospective Studies
  • Renal Dialysis*
  • Reproducibility of Results