Chip-based electrochromatography coupled to ESI-MS detection

Electrophoresis. 2016 May;37(10):1345-52. doi: 10.1002/elps.201500543. Epub 2016 Mar 15.

Abstract

In this study, we present the coupling of chip-based electrochromatography to MS using a glass chip with a monolithically integrated nanoelectrospray emitter. As separation column, an acrylate-based porous polymer monolith is implemented into the glass chip by photopolymerization. For the establishment and development of this method, we used a test mixture detectable with both fluorescence and ESI-MS. After successful evaluation of the approach with the test solutes, it was applied exemplarily for drug analysis such as high-speed separations of benzodiazepines in pharmaceuticals.

Keywords: Electrochromatography; Mass spectrometry; Microchip.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Benzodiazepines / analysis
  • Capillary Electrochromatography / methods*
  • Fluorescence
  • Lab-On-A-Chip Devices*
  • Polymers / chemistry
  • Porosity
  • Spectrometry, Mass, Electrospray Ionization / instrumentation*

Substances

  • Polymers