Optimalization of Poly(neutral red) Coated-wire Electrode for Determination of Citrate in Soft Drinks

Sensors (Basel). 2008 Feb 4;8(2):594-606. doi: 10.3390/s8020594.

Abstract

This report presents an optimization of potentiometric measurements withcitrate-selective electropolymerized poly(neutral red) electrodes. The optimal backgroundelectrolyte for these measurements is a TRIS buffer with nitrate at pH 8.5. The electrodesdescribed here exhibit stable and reproducible near-Nernstian response to citrates with alow detection limit of 6 × 10-6 M. Electrodes polymerized from sulfuric acid andacetonitrile are compared in detail. Simple and sensitive method for quantification ofcitrate in real-life samples by potentiometry with poly(neutral red) electrodes arepresented. Data from potentiometric measurements of citrate are compared with capillaryelectrophoresis.

Keywords: citrate; coated-wire electrode; conductive polymer; poly(neutral red); real-life sample.