[Assessment of the surgeon radiation exposure during a minimally invasive TLIF: Comparison between fluoroscopy and O-arm system]

Neurochirurgie. 2015 Aug;61(4):255-9. doi: 10.1016/j.neuchi.2015.04.002. Epub 2015 Jun 10.
[Article in French]

Abstract

Introduction: Transforaminal lumbar interbody fusion with a minimally invasive approach (MIS TLIF) has become a very popular technique in the treatment of degenerative diseases of the lumbar spine, as it allows a decrease in muscle iatrogenic. However, iterative radiological controls inherent to this technique are responsible for a significant increase in exposure to ionizing radiation for the surgeon. New techniques for radiological guidance (O-arm navigation-assisted) would overcome this drawback, but this remains unproven.

Objectives: To analyze the exposure of the surgeon to intraoperative X-ray during a MIS TLIF under fluoroscopy and under O-arm navigation-assisted.

Materials and methods: This prospective study was conducted at the University Hospital of Lille from February to May 2013. Twelve patients underwent a MIS TLIF for the treatment of low-grade spondylolisthesis; six under standard fluoroscopy (group 1) and six under O-arm system (group 2). Passive dosimeters (rings and glasses) and active dosimeters for thorax were used to measure the radiation exposure of the surgeon.

Results: For group 1, the average time of fluoroscopy was 3.718 minutes (3.13-4.56) while no radioscopy was perform on group 2. For the first group, the average exposure dose was 12 μSv (5-20 μSv) on the thorax, 1168 μSv (510-2790 μSv) on the main hand and 179 μSv (103-486 μSv) on the lens. The exposure dose was measured zero on the second group.

Conclusion: The maximum recommended doses can be reached, mainly for the lens. In addition to the radioprotection measures, O-arm navigation systems are safe alternatives to significantly reduce surgeon exposure.

Keywords: Arthrodèse lombaire transforaminale; Chirurgie du rachis minimale invasive; Exposition aux rayonnements; Fluoroscopie peropératoire; Intraoperative O-arm navigation-assisted; Intraoperative fluoroscopy; Minimally invasive surgery; Neuronavigation O-arm peropératoire; Radiation exposure; Radiation safety; Sûreté radiologique; Transforaminal lumbar interbody fusion.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Adult
  • Female
  • Fluoroscopy*
  • Humans
  • Lumbar Vertebrae / surgery*
  • Male
  • Middle Aged
  • Minimally Invasive Surgical Procedures / instrumentation*
  • Prospective Studies
  • Radiation Exposure*
  • Spinal Fusion / methods