Comparison of Postoperative Coronal Leg Alignment in Customized Individually Made and Conventional Total Knee Arthroplasty

J Pers Med. 2021 Jun 12;11(6):549. doi: 10.3390/jpm11060549.

Abstract

Neutral coronal leg alignment is known to be important for postoperative outcome in total knee arthroplasty (TKA). Customized individually made implants (CIM) instrumented with patient-specific cutting guides are an innovation aiming to increase the precision and reliability of implant positioning and reconstruction of leg alignment. We aimed to compare reconstruction of the hip-knee-ankle angle (HKA) of the novel CIM system iTotal™ CR G2 (ConforMIS Inc.) to a matched cohort of the off-the-shelf (OTS) knee replacement system Vanguard™ CR (Zimmer Biomet). Retrospective analysis of postoperative coronal full-leg weight-bearing radiographs of 562 TKA (283 CIM TKA, 279 OTS TKA) was conducted. Via a medical planning software, HKA and rotation of the leg were measured in postoperative radiographs. HKA was then adjusted for rotational error, and 180° ± 3° varus/valgus was defined as the target zone HKA. Corrected postoperative HKA in the CIM group was 179.0° ± 2.8° and 179.2° ± 3.1° in the OTS group (p = 0.34). The rate of outliers, outside of the ±3° target zone, was equal in both groups (32.9%). Our analysis showed that TKA using patient-specific cutting guides and implants and OTS TKA implanted with conventional instrumentation resulted in equally satisfying restoration of the coronal leg alignment with less scattering in the CIM group.

Keywords: custom-made implant; leg alignment; patient-specific instruments; rotational correction; total knee arthroplasty.