Analysis of antibiotics in urine and wipe samples from environmental and biological monitoring--comparison of HPLC with UV-, single MS- and tandem MS-detection

J Chromatogr B Analyt Technol Biomed Life Sci. 2006 Feb 2;831(1-2):72-80. doi: 10.1016/j.jchromb.2005.11.030. Epub 2005 Dec 9.

Abstract

Results of the simultaneous determination of the structurally different antibiotics cefazoline, cefotiame, cefuroxime, chloramphenicol, ciprofloxacin, ofloxacin, sulfamethoxazole and trimethoprim from environmental and biological monitoring using high-performance liquid chromatography with UV, single mass and tandem mass spectrometry were compared. For sample enrichment and clean-up a SPE method using bakerbond C18 cartridges was developed. Mean recovery rates were above 70%. Because of the complex urine matrix, only the wipe samples could be analyzed by UV-detection. However, UV-detection and single MS-detection are useful for control measurements after spillage, e.g. (LOD=1-2 ng/cm(2)). Samples from biological monitoring of occupational uptake should be analyzed by LC-MS/MS. The limits of detection (LOD) in urine ranged from 0.4 to 70 microg/L for LC-MS and 0.01 to 0.9 microg/L for LC-MS/MS detection. The limits of detection in wipe samples ranged from 0.003 to 0.13 ng/cm(2).

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Anti-Bacterial Agents / urine*
  • Chromatography, High Pressure Liquid / methods
  • Environmental Exposure / analysis*
  • Mass Spectrometry / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet

Substances

  • Anti-Bacterial Agents