Design and Analyses of a Transdermal Drug Delivery Device (TD3) †

Sensors (Basel). 2019 Nov 21;19(23):5090. doi: 10.3390/s19235090.

Abstract

In this paper, we introduce a novel type of transdermal drug delivery device (TD3) with a micro-electro-mechanical system (MEMS) design using computer-aided design (CAD) techniques as well as computational fluid dynamics (CFD) simulations regarding the fluid interaction inside the device during the actuation process. For the actuation principles of the chamber and microvalve, both thermopneumatic and piezoelectric principles are employed respectively, originating that the design perfectly integrates those principles through two different components, such as a micropump with integrated microvalves and a microneedle array. The TD3 has shown to be capable of delivering a volumetric flow of 2.92 × 10-5 cm3/s with a 6.6 Hz membrane stroke frequency. The device only needs 116 Pa to complete the suction process and 2560 Pa to complete the discharge process. A 38-microneedle array with 450 µm in length fulfills the function of permeating skin, allowing that the fluid reaches the desired destination and avoiding any possible pain during the insertion.

Keywords: computational fluid dynamic; finite element analysis; micro-electro-mechanical systems (MEMS); microstructures; transdermal drug delivery.

MeSH terms

  • Administration, Cutaneous
  • Computer-Aided Design / instrumentation
  • Drug Delivery Systems / instrumentation*
  • Equipment Design / instrumentation*
  • Humans
  • Micro-Electrical-Mechanical Systems / instrumentation*
  • Microinjections / instrumentation
  • Needles
  • Pharmaceutical Preparations / administration & dosage
  • Skin / drug effects

Substances

  • Pharmaceutical Preparations