A new automatic system for angular measurement and calibration in radiometric instruments

Sensors (Basel). 2010;10(4):3703-17. doi: 10.3390/s100403703. Epub 2010 Apr 13.

Abstract

This paper puts forward the design, construction and testing of a new automatic system for angular-response measurement and calibration in radiometric instruments. Its main characteristics include precision, speed, resolution, noise immunity, easy programming and operation. The developed system calculates the cosine error of the radiometer under test by means of a virtual instrument, from the measures it takes and through a mathematical procedure, thus allowing correcting the radiometer with the aim of preventing cosine error in its measurements.

Keywords: angular response; calibration; cosine error; data acquisition; radiometric instrument; solar irradiance.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / methods*
  • Calibration
  • Databases, Factual
  • Equipment Design / methods*
  • Humans
  • Radiometry / instrumentation*