Characterization of a 0.8 mm3Medscint plastic scintillator detector system for small field dosimetry

Phys Med Biol. 2023 Aug 29;68(17). doi: 10.1088/1361-6560/aceacf.

Abstract

Objective. Plastic scintillator detectors (PSDs) have demonstrated ability to meet requirements of small field dosimetry. Medscint developed a 1 mm long, 1 mm diameter cylindrical PSD with effective volume of 0.8 mm3. Clinically relevant, small field dosimetric properties of this detector, combined with a novel scintillation dosimetry system-HYPERSCINT RP-200, and HYPERDOSE analysis software were evaluated in this study.Approach. This novel scintillator-based dosimetry system was characterized with 6 MV-WFF and 10 MV-FFF x-ray beams delivered by Varian TrueBeamTMlinear accelerator. The detector was characterized for leakage, short-term repeatability, dose response linearity, angular response, dose rate response, and field size dependence for radiation field sizes of 0.25 × 0.25 to 10 × 10 cm2. Measured detector specific output ratios were compared with microDiamond output factors to determine small field output correction factors,kQclin,Qmsrfclin,fmsr.Main results. The dosimetry system showed excellent short-term repeatability with standard deviation of only 0.04 ± 0.01%. It demonstrated good dose linearity with variations less than 1.0% for 14.4 cGy and above. The dosimetry system was found to be independent of dose rate and angle of irradiation, with deviations for both below 0.5%. Leakage was found to be comparable to background readings. For 6 MV-WFF energy beams, detector specific output ratios for field sizes down to 1 × 1 cm2agreed with output factors measured with PTW TN60019 microDiamond, thus,kQclin,Qmsrfclin,fmsrequates to unity for these field sizes. For 10 MV-FFF energy beams, detector specific output ratios for field sizes down to 2 × 2 cm2agreed with PTW TN60019 microDiamond output factors, thus,kQclin,Qmsrfclin,fmsrequates to unity for these field sizes.kQclin,Qmsrfclin,fmsrfor field sizes down to 0.5 × 0.5 cm2were determined to be within 6% of unity for both 6 MV-WFF and 10 MV-FFF energy beams.Significance. The HYPERSCINT RP-200 dosimetry system coupled with a 0.8 mm3PSD showed excellent dosimetric properties and was found to be clinically relevant for relative dosimetry down to field sizes of 0.5 × 0.5 cm2and potentially smaller.

Keywords: plastic scintillator detector; scintillation dosimetry; small field dosimetry.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Monte Carlo Method
  • Particle Accelerators*
  • Photons
  • Radiometry* / methods
  • Software
  • X-Rays