Assessing Spatiotemporal and Quality Alterations in Paretic Upper Limb Movements after Stroke in Routine Care: Proposal and Validation of a Protocol Using IMUs versus MoCap

Sensors (Basel). 2023 Aug 25;23(17):7427. doi: 10.3390/s23177427.

Abstract

Accurate assessment of upper-limb movement alterations is a key component of post-stroke follow-up. Motion capture (MoCap) is the gold standard for assessment even in clinical conditions, but it requires a laboratory setting with a relatively complex implementation. Alternatively, inertial measurement units (IMUs) are the subject of growing interest, but their accuracy remains to be challenged. This study aims to assess the minimal detectable change (MDC) between spatiotemporal and quality variables obtained from these IMUs and MoCap, based on a specific protocol of IMU calibration and measurement and on data processing using the dead reckoning method. We also studied the influence of each data processing step on the level of between-system MDC. Fifteen post-stroke hemiparetic subjects performed reach or grasp tasks. The MDC for the movement time, index of curvature, smoothness (studied through the number of submovements), and trunk contribution was equal to 10.83%, 3.62%, 39.62%, and 25.11%, respectively. All calibration and data processing steps played a significant role in increasing the agreement. The between-system MDC values were found to be lower or comparable to the between-session MDC values obtained with MoCap, meaning that our results provide strong evidence that using IMUs with the proposed calibration and processing steps can successfully and accurately assess upper-limb movement alterations after stroke in clinical routine care conditions.

Keywords: inertial sensor; kinematics; motion capture; stroke rehabilitation; upper extremity.

MeSH terms

  • Blood Coagulation Tests
  • Humans
  • Laboratories
  • Motion Capture*
  • Movement
  • Stroke*

Substances

  • ethoprop

Grants and funding

This research received no external funding.