Abstract

The article presents the influence of the camera and its optical system on the uncertainty of object position measurement in vision systems. The aim of the article is to present the methodology for estimating the combined standard uncertainty of measuring the object position with a vision camera treated as a measuring device. The identification of factors affecting the location measurement uncertainty and the determination of their share in the combined standard uncertainty will allow determining the parameters of the camera operation, so that the expanded uncertainty is as small as possible in the given measurement conditions. The analysis of the uncertainty estimation presented in the article was performed with the assumption that there is no influence of any external factors (e.g., temperature, humidity, or vibrations).