Intraoperative Assessment of High-Risk Thyroid Nodules Based on Electrical Impedance Measurements: A Feasibility Study

Diagnostics (Basel). 2022 Nov 25;12(12):2950. doi: 10.3390/diagnostics12122950.

Abstract

Precise diagnosis of thyroid nodules is challenging due to non-diagnostic/inconclusive results and uncertainties about the malignancy of follicular neoplasms (FNs), even in frozen-section pathology. Therefore, surgical management, especially in Bethesda III and IV categories, may be complicated, and sometimes a second surgery may be required. The Thyroid Nodule Impedance Measurement System (TN-IMS) consists of a metallic patch attached to submental skin and a G20 I.V. cannula inserted into the targeted nodules. Two impedance-based parameters named Z1kHz and impedance phase slope (IPS) in 100 kHz to 500 kHz of the thyroid nodules are recorded and compared with their histopathological results as the gold standard. TN-IMS was intra-surgically applied to 103 human thyroid nodules and normal thyroid tissues. A remarkable consistency between defined co-ranges of Z1kHz/IPS and the histopathological status of specimens was achieved (p < 0.001). Based on these measurements, it was concluded that intraoperative bioelectrical impedance scanning of thyroid nodules would be a helpful complementary approach to detecting high-risk excision-required thyroid nodules.

Keywords: atypia of undetermined significance (AUS); electrical impedance spectroscopy; fine needle aspiration; impedance phase slope (IPS); surgery; thyroid nodule.

Grants and funding

This research received no external funding.