The non-contact heart rate measurement system for monitoring HRV

Annu Int Conf IEEE Eng Med Biol Soc. 2013:2013:3238-41. doi: 10.1109/EMBC.2013.6610231.

Abstract

A noncontact ECG monitoring and analysis system was developed using capacitive-coupled device integrated to a home sofa. Electrodes were placed on the backrest of a sofa separated from the body with only the chair covering and the user's clothing. The study also incorporates measurements using different fabric materials, and a pure cotton material was chosen to cover the chair's backrest. The material was chosen to improve the signal to noise ratio. The system is initially implemented on a home sofa and is able to measure non-contact ECG through thin cotton clothing and perform heart rate analysis to calculate the heart rate variability (HRV) parameters. It was also tested under different conditions and results from reading and sleeping exhibited a stable ECG. Subsequently, results from our calculated HRV were found to be identical to those of a commercially available HRV analyzer. However, HRV parameters are easily affected by motion artifacts generated during drinking or eating with the latter producing a more severe disturbance. Lastly, parameters measured are saved on a cloud database, providing users with a long-term monitoring and recording for physiological information.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Drinking
  • Eating
  • Electrocardiography / instrumentation
  • Electrocardiography / methods*
  • Electrodes
  • Heart Rate / physiology*
  • Humans
  • Monitoring, Physiologic
  • Reading
  • Sleep