A study on the correlation between patients' physical characteristics and effective dose of liver computed tomography

Technol Health Care. 2014;22(3):403-8. doi: 10.3233/THC-140797.

Abstract

This suggests indicators to be considered in the protocol for setting up equipment and minimizing patient doses by identifying the effective dose and correlations of each equipment company according to a patient's body characteristics in liver CT. The study was conducted with 445 patients who went to the hospital and received liver CT at the diagnostic radiology department of S medical center from 2010 January to June. As the statistical methods, t-test, one-way ANOVA, and Pearson's correlation analysis were used. The study results show that as height, weight, and BMI increased, the effective dose increased with all equipment vendors. Correlations between a patient's body characteristics and the effective dose were shown to be positive with all equipment vendors in regard to weight, BMI, and height, in order.

Keywords: Body Mass Index (BMI); Liver CT; correlation; effective dose.