Simple high-performance liquid chromatographic method for the determination of ranitidine in human plasma

J Chromatogr B Biomed Sci Appl. 1998 Oct 23;718(1):205-10. doi: 10.1016/s0378-4347(98)00359-4.

Abstract

A simple high-performance liquid chromatographic method was developed for the determination of ranitidine in human plasma. Prior to analysis, ranitidine and the internal standard (metoprolol) were extracted from alkalinized plasma samples using dichloromethane. The mobile phase was 0.05 M potassium dihydrogenphosphate-acetonitrile (88:12, v/v) adjusted to pH 6.5. Analysis was run at a flow-rate of 1.3 ml/min and at a detection wavelength of 229 nm. The method is sensitive with a detection limit of 1 ng/ml at a signal-to-noise ratio of 3:1, while the quantification limit was set at 15 ng/ml. The calibration curve was linear over a concentration range of 15-2000 ng/ml. Mean recovery value of the extraction procedure was about 90%, while the within-day and between-day coefficients of variation and percent error values of the assay method were all less than 15%.

MeSH terms

  • Adult
  • Biological Availability
  • Chromatography, High Pressure Liquid / methods*
  • Evaluation Studies as Topic
  • Humans
  • Male
  • Ranitidine / blood*
  • Ranitidine / pharmacokinetics
  • Sensitivity and Specificity

Substances

  • Ranitidine